ClearanceBrief
Recall Z-1226-2022 ยท CBK Ventilator, Continuous, Facility Use

GE Healthcare, LLC

Back up batteries in Ventilators labeled as the following: a. CARESCAPE R860 Ventilators (K142679) b. ENGSTROM (K111116) c. ENGSTROM Carestation (K111116) d. ENGSTROM PRO (K111116)

DateJun 6, 2022 (initiated Apr 18, 2022)
StatusOpen, Classified
ReasonInsufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.
Root causeUnder Investigation by firm
ActionThe firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter the consignees via mail on 04/18/2022. The letter describes the product, problem and actions to be taken. The customers instructed to take the following actions: 1. You can continue to use the affected ventilators while the ventilator is connected to an AC mains power source that is supported by backup emergency power. 2. If it is absolutely necessary to use the ventilator by relying on the battery (such as during required transport where there are no other options), ensure you follow standard clinical practice of
Quantity34,940 ventilators
DistributionWorldwide Distribution: US (nationwide) to AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and OUS (internationally) to Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Chi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111116ENGSTROM CARESTATION, ENGSTROM PRO, MODELS 1505-9000-000, 1505-9003-000
Datex-Ohmeda, Sep 9, 2011
K142679CARESCAPE R860
Datex-Ohmeda, Inc., Jun 12, 2015