Recall Z-1216-2022 ยท LXH Orthopedic Manual Surgical Instrument

Tornier, Inc

Ratcheting Screwdriver Handle, provided in the Aequalis Perform Reversed kit

DateMay 3, 2022
StatusOpen, Classified
ReasonThe device is unable to meet sterilization requirements when sterilized per instructions in an Aesculap container.
Root causeDevice Design
ActionThe recalling firm issued recall notification letters on May 3, 2022 via UPS ground. Consignees are asked to do the following: 1. Immediately cease sterilizing the kitted Aequalis" Perform" Reversed Ratcheting Screwdriver Handle (MWJ128) in an Aesculap container. a. If sterilization in an Aesculap container is desired, the handle (MWJ128) shall be removed from the kit (YKAD261) prior to sterilization and then sterilized separately. To ensure sterility, the handle (MWJ128) must be double-wrapped and sterilized individually according to the parameters provided in the table below and provided w
Quantity1066 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161742Aequalis PerFORM Reversed, Aequalis PerFORM+ Reversed Glenoid
Tornier, Inc., Nov 15, 2016