Recall Z-1097-2022 ยท CGS Nad Reduction/Nadh Oxidation, Cpk Or Isoenzymes

Randox Laboratories Ltd.

CK MB Calibrator, Catalogue Number CK2393

DateMar 25, 2022
StatusTerminated
ReasonRandox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples.
Root causeUnder Investigation by firm
ActionFirm notified customers via email on March 25, 2022. The distributors and customers were instructed to discard previous target sheet and perform calibration with the realigned target value provided with the letter. Also, to complete and return the response form 12187-QA to technical.service@randox.com.
Quantity1,392
DistributionDomestic distribution to WV and PR. International distribution to Bangladesh, Bosnia & Herzegovina, Botswana, China, Colombia, Costa Rica, Croatia, Egypt, Fiji, France, Georgia, Germany, Ghana, Greece, Grenada, Hong Kong, India, Iraq, Italy, Lithuania, Malaysia, Maldives, Malta, Mexico, Morocco, Moz
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.