Recall Z-1208-2022 ยท GKA Abnormal Hemoglobin Quantitation

Bio-Rad Laboratories, Inc.

Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reorder Pack, REF: 250-3000

DateApr 7, 2022
StatusOpen, Classified
ReasonThe problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in specific lots of VARIANT nbs Sickle Cell Progam Reorder Pack causes all customized settings in the Setup/Test/Data Setup subscreen and Setup/Test/Pattern Setting subscreen to be overwritten with the default settings when the Update Kit procedure is performed.The VARIANTnbs Sickle Cell Program has been validated using the default Resin Update CDROM parameters for the intended use.
Root causePackaging process control
ActionOn 04/07/22, Field Action Notifications were sent to customers, who were informed of the following. Not affected by this action if A) affected Resin Update CD-ROM was not installed. B) Default settings are used (i.e., did not edit any settings) in the Setup/Test/Data Setup and/or Pattern Setting subscreens prior to performing the Update Kit procedure using one of the affected CD-ROMs. Affected by this action if customized settings were used and subsequently the Update Kit procedure using one of the affected CD-ROMs was performed; your customized settings were overwritten by the default Bio-Rad
Quantity1893
DistributionUS: AZ, AR, GA, CO, TN, LA, VA, FL, KY, IN, KS, MD, MN, MO, NC, NJ, NY, OH, OR, SC, IA, AL, UT, PR OUS: Canada, Spain, France, Italy, India, Vietnam, Philippines
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051072BIO-RAD VARIANT NBS SICKLE CELL PROGRAM
Bio-Rad Laboratories, Inc., May 12, 2005