Recall Z-1092-2022 ยท OOI Real Time Nucleic Acid Amplification System

Luminex Corporation

Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system

DateMay 13, 2022 (initiated Apr 5, 2022)
StatusOpen, Classified
ReasonHeater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.
Root causeUnder Investigation by firm
ActionOn 04/21/22 Luminex issued a "Urgent Medical Device Recall" Notification to affected consignees. On 6/16/22 and 7/14/22 Luminex expanded the recall to include additional serial numbers. Luminex is asking consignees to take the following actions: 1. Continue use of your VERIGENE Processor SP as service has been completed by a Luminex Field Service Engineer and the Processor SP s heaters are now within specification. 2. Perform a retrospective review of results for VERIGENE Gram-Positive Blood Culture (BC-GP) and VERIGENE Gram-Negative Blood Culture (BC-GN) Nucleic Acid Tests from 18 months
Quantity24 units
DistributionUS:OH, ME, PA, FL, IL, NE, IN, TX, GA, WA, MI, CO, AZ, OH, KY, CA, VT OUS: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143653Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex)
Nanosphere, Inc., Sep 4, 2015