Recall Z-1052-2022 ยท GJS Test, Time, Prothrombin

Instrumentation Laboratory

HemosIL ReadiPlasTin, Part No. 0020301400

DateApr 8, 2022
StatusOpen, Classified
ReasonVariable and out of specification QC results with HemosIL ReadiPlasTin. IL will remove all lots from the market and convert customers to an alternative prothrombin time (PT) reagent, specifically HemosIL RecombiPlasTin 2G.
Root causeUnder Investigation by firm
ActionOn April 8, 2022, the firm notified customers via email and certified mail. Customers were informed that the firm will stop shipment of product on May 9, 2022. After receipt of the material, customers must complete the conversion to HemosIL RecombiPlasTin 2G, with support as needed from your local Werfen representative, within 90 days. Until a lot of HemosIL RecombiPlasTin 2G has been received and ready for use, customers should run quality control for HemosIL ReadiPlasTin with each new reagent vial (every lot) and at least every 8 hours to identify vials with performance issues. Any product
Quantity1,563 (US); 29,046 (OUS)
DistributionUS Nationwide distribution including in the states of AZ, CA, IA, IL, IN, KS, MD, ME, MI, MN, MO, NC, NE, NY, OH, OK, SC, TX, VA, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122584HEMOSIL READIPLAS TIN
Instrumentation Laboratory CO, Mar 24, 2013