Recall Z-1123-2022 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Medtronic Neuromodulation

Vanta INS Implantable Neurostimulator. Medtronic Model 977006 Vanta with AdaptiveStim Technology Neurostimulator with Surescan MRI technology, part of a neurostimulator system for pain therapy.

DateMay 23, 2022 (initiated Mar 30, 2022)
StatusOpen, Classified
ReasonThe Implantable Neurostimulator Model may be susceptible to not communicating with the clinical programmer and/or patient programmer system.
Root causeUnder Investigation by firm
ActionA customer letter, dated March 2022, was issued to the impacted consignee via mail. The consignee is instructed to ensure the patient programmer system is paired with the patient's INS. If the INS cannot communicate with the clinician programmer and/or patient programmer system, and it is not possible to make therapy adjustments, the consignee is to contact Medtronic Technical Services. The consignee is also asked to complete the customer confirmation form and submit to the recalling firm. Medtronic will be notifying consignees and actively retrieving any unused Vanta INS devices that con
Quantity1021 devices
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Italy, Sweden,