Recall Z-1094-2022 ยท ILQ System, Communication, Powered

Baxter Healthcare Corporation

Voalte Nurse Call System provides a comprehensive communication and information system that places patient calls, staff calls, and emergency calls.

DateApr 13, 2022
StatusTerminated
ReasonFirm discovered a firmware memory leak with a supplier-manufactured component.
Root causeSoftware design
ActionThe firm, Hillrom, issued an "Urgent Medical Device Correction Notice", dated 04/13/2022, to impacted consignees via email. The consignee is instructed to identify if they have the impacted product component, Part P2519NNC3A2 and also distribute the notification to personnel as well ensure that the users monitor for system alerts. Notify Hillrom Technical support if there is a loss of connectivity to the nurse call server. Note, error indications will display on multiple user interfaces. Central nurses' stations will state the room(s) is down (disconnected). In patient room(s), graphical au
Quantity1414 units
DistributionUS Distribution to states of: NC and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.