Recall Z-1098-2022 ยท FSA Lift, Patient, Non-Ac-Powered

ARJOHUNTLEIGH POLSKA Sp. z.o.o.

Sara Plus Active Floor Lift

DateMay 21, 2022 (initiated Apr 5, 2022)
StatusTerminated
ReasonThe device may emit smoke or ignite.
Root causeDevice Design
ActionA customer letter was sent to consignees on 04/08/2022. The customer are asked to acknowledge the notification, complete and sign off an enclosed Customer Response Form and return it to the local Arjo office (address indicated in the notification). The notification contains the list of precautionary measures to be taken by caregivers in order to minimize the likelihood of the failure until their lifts are corrected by Arjo. The device correction (including PCB upgrade and cover replacement) is to be performed by Arjo personnel, free of charge.
Quantity4449 devices
DistributionUS Nationwide and Worldwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.