Recall Z-1021-2022 ยท CHP Radioimmunoassay, Estradiol

Siemens Healthcare Diagnostics, Inc.

Atellica IM Enhanced Estradiol (eE2) 500T - In vitro diagnostic use in the quantitative determination of estradiol SMN: 10995562 All lots

DateMar 3, 2022
StatusOpen, Classified
ReasonFalsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Root causeUnder Investigation by firm
ActionSiemens issued an Urgent Medical Device Correction AIMC 22-03.A.US and Urgent Field Safety Notice AIMC 22-03.A.OUS on 3/3/22. Letter states reason, health risk and action to take: Customers must discontinue use of plasma specimens for testing with the Atellica IM and ADVIA Centaur eE2 assays until further notice. Customers may continue to use the Atellica IM eE2 and ADVIA Centaur eE2 assays with serum specimens. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any
Quantity374 units
DistributionWorldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Bulgaria, Burkina, Faso Country AG, Cyprus, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Georgia, Greece, Guadeloupe, India, Ireland, Israel, Italy, Kazakhstan, Latvia, Lithuania, Macedonia, Malta, M
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.