ClearanceBrief
Recall Z-1048-2022 ยท EOQ Bronchoscope (Flexible Or Rigid)

Olympus Corporation of the Americas

BF-MP160F: EVIS EXERA Bronchofibervideoscope

DateMar 8, 2022
StatusOpen, Classified
ReasonBronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using sterilization
Root causeOther
ActionOlympus issued Urgent Medical Device Correction Letters on 3/8/22. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory for the referenced devices and identify any device with the model names specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Carefully read the content of this Medical Device Correction Action as well as the attached Addendum . The addendum provides updated information or instructions on the chemicals to be used for reprocessing, the sterilization conditions, rinsing after
Quantity292 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033225EVIS EXERA BRONCHOFIBERVIDEOSCOPE OLYMPUS BF TYPE XP160F, ITS ACCESSORIES AND ANCILLARY EQUIPMENT
Olympus Optical Co., Ltd., Oct 16, 2003