Recall Z-0994-2022 ยท MUJ System, Planning, Radiation Therapy Treatment
Siemens Medical Solutions USA, Inc
syngo.via RT Image Suite upgraded to VB60 running on syngo.via Model Number 10496180. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
| Date | Mar 7, 2022 |
| Status | Terminated |
| Reason | Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step. |
| Root cause | Software change control |
| Action | 4/4/22 update: Urgent Medical Device Correction letter has not been issued. The proposed letter informs customers of software issue and workaround to resolve issue. It also requires electronic signature to acknowledge receipt of the notification. If you have additional questions contact the service organization at 1-800-888-7436. |
| Quantity | 1 (US) |
| Distribution | Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, IL, IN, KY, MD, MO, ND, NH, NY, OH, OR, PA, SD, TX, VA and the countries of Argentina, Austria, Bahrain, Belgium, Brazil, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Netherlands, Norway, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K211379 | syngo.via RT Image Suite Siemens Medical Solutions USA, Inc., Jul 30, 2021 |