Recall Z-0952-2022 ยท OWD Somatic Gene Mutation Detection System

Qiagen Sciences LLC

therascreen KRAS RGQ PCR Kit (24) REF 874052 (Japan IVD, not released in the USA)

DateFeb 25, 2022
StatusOpen, Classified
ReasonFalse positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutation target of the kit.
Root causeDevice Design
ActionQiagen issued Urgent Medical Device Correction therascreen KRAS RGQ PCR Kit (24), REF 870021 letter on 2/25/22. Letter states reason for recall, health risk and action to take: This notice is only relevant for the analysis of samples originating from non-small cell lung cancer (NSCLC) tissue. For continued use of the therascreen KRAS RGQ PCR Kit (24), check the Flags/Warnings in the Sample Result table for automated interpretation of results. Sample results with one or more of the Flags/Warnings listed in this notice should be regarded as invalid with respect to the KRAS status and KRAS Mutati
Quantity8 Kits
DistributionAZ CA CT FL GA IL IN MA MN NC NY OK OR PA TN TX WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.