Recall Z-0962-2022 · JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SIEMENS SOMATOM with software syngo CT VA30A_SP4 Models: SOMATOM go.Now 11061610 & 11061618 SOMATOM go.Up 11061620 & 11061628 SOMATOM go.All 11061630 & 11061638 SOMATOM go.Top 11061640 & 11061648 SOMATOM go.Sim 11061660 & 11061668 SOMATOM go.Open Pro 11061670 & 11061678 SOMATOM X.cite 11330001 Intended Use: Computed tomography x-ray system

DateMar 7, 2022
StatusOpen, Classified
ReasonSoftware error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows. Results in potential patient issues: Possible rescan , Unexpected X-Ray dose and additional contrast media, Delay in diagnosis, scan abort,and patient rescan
Root causeSoftware design
ActionSiemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice to customers via CT010/22/S via email/mail on March 7, 2022. Letter states reason for recall, health risk and action to take: There are no workarounds for these issues. Customers are urged to install the remote update as soon as it is released. How will the corrective action be implemented and what is the efficiency of the corrective actions? Siemens has developed software update syngo CT VA30A_SP5 to ensure more stable scanning workflows, reduce the number of user notifications, and to further improve the image
Quantity547 units US
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163296SOMATOM go.Up, SOMATOM go.Now
Siemens Medical Solutions USA, Inc., Mar 21, 2017
K173632SOMATOM go.Up, SOMATOM go.Now, SOMATOM go.Top, SOMATOM go.All, Scan&GO
Siemens Medical Solutions USA, Inc., Apr 13, 2018
K192061SOMATOM Go Platform – SOMATOM go.Now, go.Up, go.All, go.Top, go.Sim, go.Open Pro, Scan & GO
Siemens Medical Solutions USA, Inc., Nov 21, 2019
K200524SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All, go. Sim, go. Open Pro, Scan&GO
Siemens Medical Solutions USA, Inc., Apr 1, 2020
K211373SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X. Platform Scanners - SOMATOM X.cite and
Siemens Medical Solutions USA, Inc., Aug 27, 2021