Recall Z-1015-2022 ยท DHA System, Test, Human Chorionic Gonadotropin

Siemens Healthcare Diagnostics, Inc.

IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number 10381194, catalog number L2KCG6]

DateFeb 22, 2022
StatusOpen, Classified
ReasonPotential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.
Root causeUnder Investigation by firm
ActionUrgent Medical Device Correction was issued February 02, 2022 via Fed Ex. Notification provides instructions to customers to mitigate this sample carryover. Customers are requested to complete and return the Effectiveness Check questionnaire. Contact your local Siemens Healthineers technical support representative with any questions.
Quantity10,755 boxes (533 US, 10,222 OUS)
DistributionWorldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, ID, IN, MN, NC, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TX, VA, WI, WY and OUS: 81 countries.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K990222IMMULITE HCG, MODEL LKCG1, LCCG5, IMMULITE 2000 HCG, MODEL L2KCG2, L2KCG6
Diagnostic Products Corp., Feb 26, 1999