Recall Z-1055-2022 · EOQ Bronchoscope (Flexible Or Rigid)

Karl Storz Endoscopy

11005BC1 11005BCK1 Flexible Bronchoscope Z21493US-BA (06/2019)

DateApr 1, 2022
StatusOpen, Classified
ReasonFailure to achieve the expected six-log reduction in microorganisms following the disinfection process.
Root causeUnder Investigation by firm
ActionOn 04/01 and 6/2022, the firm emailed an "Urgent Medical Device Recall" letter to customers and followed up with a hardcopy letter on 04/07-08/2022 via their Third Party contractor informing customers of a labeling update to correct the instructions for use for certain flexible urological endoscopes and advise that high-level disinfection should not be used as a method of reprocessing. Supplemental validation testing of the efficacy of the manual high-level disinfection process using CIDEX OPA was performed. Testing showed that the required efficacy level of disinfection using CIDEX OPA was no
QuantityAll Serial Numbers manufactured/distributed since January 2018
DistributionU.S. Nationwide distribution in the states of FL, KY, LA, MA, MI, MO, MT, NC, NH, NJ, OK, SC, TX and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212656Flexible Intubation Video Endoscope – Sterile (FIVE-S)
Karl Storz Endoscopy America, Dec 29, 2021