Recall Z-1041-2022 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Philips North America Llc

Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient hemodynamic data in the Philips Interventional Hemodynamic System.

DateFeb 22, 2022
StatusTerminated
ReasonWhen a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.
Root causeUnder Investigation by firm
ActionAn URGENT Medical Device Correction notification letter dated 2/18/22 was sent to customers. Actions that should be taken by the customer/ user in order to prevent risks for patients. Do not use the Non-Invasive Blood Pressure for Neonatal Patients until Philips has installed a new software revision that will solve this problem. Place this Urgent Medical Device Correction letter with the documentation of the Philips Hemodynamic Application until Philips implements the new software revision. Return the attached reply form to Philips to confirm that the users of the system have reviewed and un
Quantity77
DistributionDevices were distributed to the following US states: FL, ID, NJ, OK, SC and TX. Devices were distributed to the following foreign countries: France, Germany, India, Mexico, Netherlands, New Zealand, Norway, Poland, South Africa, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181311Philips Hemodynamic Application R1.0
Philips Medical Systems Nederlands B.V., Sep 7, 2018