Recall Z-1040-2022 ยท OES Cardiac Catheterization Kit

American Contract Systems, Inc.

ACS Cath Lab Pack

DateMar 22, 2022
StatusOpen, Classified
ReasonProduct was sterilized with a higher than specification EO concentration.
Root causeProcess control
ActionA customer letter dated 04/05/2022 was issued to impacted consignees. The letter instructs the following: 1) Review your inventory and segregate affected product. This product should not be shipped out to customers. 2) Return affected product to 2610 NE Industrial Drive, Suite 220, Kansas City, MO 64120. 3) If product was already further distributed, ensure that all downstream customers are properly notified of the recall and maintain records of effectiveness. A Field Action Response Form should be completed and returned to suzanne.thomas@hyh.com, even if you do not have product inventory.
Quantity10 packs
DistributionUS Nationwide distribution in the states of IL, MO, NE, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.