ClearanceBrief
Recall Z-1002-2022 ยท OVB Hydrogel Spacer

Boston Scientific Corporation

SpaceOAR System, UPN SO-1010, SO-2101, SO-3101, and SO-4101. Used to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer to reduce the radiation dose delivered to the anterior rectum.

DateFeb 24, 2022
StatusOpen, Classified
ReasonUsers need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.
Root causeLabeling design
ActionAn URGENT MEDICAL DEVICE PRODUCT ADVISORY dated 2/24/22 was sent to customers. Inadvertent placement of SpaceOAR gel into a blood vessel and subsequent migration of the hydrogel outside of the pelvis, led to embolism. Boston Scientific will incorporate best practices regarding hydrogel placement during all SpaceOAR procedures into the IFU. The updated IFUs will be packaged with SpaceOAR and SpaceOAR Vue Systems after regulatory approvals are obtained for the IFU updates. This Product Advisory contains the recommended IFU updates, located in Appendix 1, intended to: -Raise awareness of em
Quantity61,852 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181465SpaceOAR System
Augmenix, Inc., Jun 25, 2018
K202224SpaceOAR System
Boston Scientific Corporation, Aug 28, 2020