Recall Z-0961-2022 ยท ERL Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece

Medtronic Xomed, Inc.

Software 1898072 IPC upgrade v 2.7.3.0, IPC System: a) CONSOLE 1898001 IPC, Product Number 1898001; b) CONSOLE IPC REFURB, Product Number EC300RF; c) MASTER CONSOLE EC300 LEGEND EHS, Product Number EC300

DateDec 15, 2021
StatusTerminated
ReasonAs a result of Medtronic's integrated power console (authentication number: 222ACBZX00018000) being used with software version: v2.7.3.0, it was discovered that there was a defect that resulted in a difference between the set number of rotations and the actual number of rotations in a specific mode when the product was used with our company's M5 handpiece.
Root causeUnder Investigation by firm
ActionThe firm notified the Japanese consignees by letter on 12/12/2021. The letter stated the following: "In our company's records, the software version of your hospital's integrated power console (IPC) is not v2.7.3.0. However, we found that there were cases in other facilities that the software version was updated to v2.7.3.0 was not recorded in our company's records even though the software version was updated to v2.7.3.0.. Therefore, we proceed the following measures. Thank you for your understanding and cooperation. Content l If it has been confirmed that the software used in your hospital
Quantity1447 units
DistributionInternational distribution country of Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081277XPS 4000 SYSTEM, MIDAS REX LEGEND EHS SYSTEM, INTEGRATED POWER CONS0LE (IPC)
Medtronic Xomed, Inc., Sep 5, 2008
K081475ELECTRIC DRILL SYSTEM [XPS 4000 SYSTEM, MIDAS REX LEGEND EHS SYSTEM, INTEGRATED POWER CONSOLE (IPC)]
Medtronic Xomed, Inc., Oct 17, 2008