Recall Z-1013-2022 ยท KDN System, Perfusion, Kidney

Waters Medical Systems LLC

WAVES WCM Kidney Cassette Module

DateMar 21, 2022
StatusTerminated
ReasonThere is a potential for the WCM cassette pump head leaking solution.
Root causeProcess control
ActionWMS communicated the recall through a recall letter on 03/15/2022. The letter instructed all customers to return the product to WMS for evaluation. The goal is 100% product location and recovery. In addition, the letter instructed all customers to fill out information on the recall notification and return to WMS. For like product not involved in the recall, a warning letter was issued. For any product not included in the recall but found to be deficient, customers will be instructed to contact WMS for replacement.
Quantity75 cases (6ct)
DistributionWCM cassettes have been distributed to one domestic customer with two locations in Tennessee and one international distributor located in France.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111521WAVES
Waters Medical Systems, LLC, Feb 16, 2012