Recall Z-0988-2022 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

St. Jude Medical, Cardiac Rhythm Management Division

Gallant DR Implantable Cardioverter Defibrillator REF # CDDRA500Q

DateMar 10, 2022
StatusOpen, Classified
ReasonBluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.
Root causeComponent design/selection
ActionAbbott issued a Safety Notification For a subset of Gallant ICDS and CRT-D, Models CDVRA500Q, CDDRA500Q and CDHFA500Q on March 10, 2022. Notification was hand delivered in US. The notification includes the Patient Management Recommendations: -As the devices are unable to be followed using remote monitoring via Bluetooth connectivity, follow-ups should be conducted in-clinic using inductive telemetry. - Abbott has developed a process to attempt recovery of Bluetooth communication via a device firmware upgrade. If remote monitoring via Bluetooth communication is desired, please work with your
Quantity1
DistributionWorldwide distribution - US Nationwide distribution in the states of FL, KY and the country of France, Czech Republic.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.