Recall Z-0904-2022 ยท DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
Merit Medical Systems, Inc.
AccessPLUS Large Bore Hemostasis Valves, Catalog Code: 580758001, 580774001, IN8152/C, K04-01197, K05T-01863, K09-09102B, K09-10388A, K09-12917, K09-13501AP, K10-05457AP, K10-05635, K10-05868P, K10-05872P, K12-08303, K12-10346, K12-10375, K12-10724AP, K12-10962, K12-11413, MAP150/F, MAP152/B, MAP152/F, K12-01551, K12-YP064A, K12-YP114, K12-YP115, K12-YS047A, IN8152, K12-01551
| Date | Mar 1, 2022 |
| Status | Open, Classified |
| Reason | A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire through the device. This defect may result in a delay of procedure. |
| Root cause | Component change control |
| Action | On 03/01/22, recall notices were mailed to customers wo were informed to take the following actions: 1. Quarantine and discontinue use, and distribution of affected devices. 2. Ensure that applicable personnel within their organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure the recall notice is immediately shared with them and note the quantity distributed on the Customer Response Form. 4. Fill out, scan, and email the completed Customer Response Form to Customer Service at response@merit. |
| Quantity | 58963 |
| Distribution | US: MN, PA, AZ, FL, IL, AK, NY, CA, TX, KY, WI, NC, NJ, VA, CO, IN, AR, MO, LA, OR, MT, MA, MI, NE, WV, KS, MS, HI, IA, NH, ME, AL, DC, DE, NM, OH, GA, MD, TN, UT, RI, SC, OK, WA, ID, ND, SD, NV, VT. OUS: Poland, Germany, France, United Kingdom, Belgium, Korea, Republic of, Austria, Canada, Spain, |
| FDA record | Open on fda.gov |