ClearanceBrief
Recall Z-0964-2022 ยท OSR Pacemaker/Icd/Crt Non-Implanted Components

St. Jude Medical, Cardiac Rhythm Management Division

Merlin PCS 3650 programmer Model 3330 software

DateMar 10, 2022
StatusOpen, Classified
ReasonDue to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters.
Root causeSoftware design
ActionThe Recalling firm is sending an "Urgent Medical Device Correction" Letter to physicians and healthcare professionals informing them of a programmer software anomaly that may be encountered when executing a pacing capture Decrement Test in-clinic on a Gallant, or Entrant device using a Merlin Patient Care System (PCS) programmer. If a user presses the Hold to Test button and decides to stop testing by releasing the button prior to the first voltage decrement (within a narrow time window approximately 2.5 seconds after test initiation), the programmer may continue to execute the Decrement Tes
Quantity30,069 devices with software
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN. IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI and WY. The countries of Argentina
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.