ClearanceBrief
Recall Z-0925-2022 ยท MAV Syringe, Balloon Inflation

Abbott Vascular

Plus 30 Priority Pack w/.096 RHV - REF 1000183 WPL2122277-01 (20-03-01) Plus 30 Priority Pack w/.115 RHV - REF 1000185-115 WPL2122279-01 (2019-03-01)

DateMar 11, 2022
StatusOpen, Classified
ReasonDue to an increase in complaint trend for leaks and intermittent/loose connections.
Root causeComponent change control
ActionOn 03/11/2022, the firm sent an "URGENT FIELD SAFETY NOTICE/DEVICE RECALL" Letter via visit, email or mail, to customers informing them that specific lots of 20/30 INDEFLATOR, INDEFLATOR Plus 30 and associated Priority Packs my exhibit leaks and/or a loose connection at the rotating luer assembly or stopcock connection, which could lead to air ingress under vacuum. Customers are instructed to: 1)Immediately stop using the devices from the affected lots 2) Review their inventory, complete and return the provided Effectiveness Check Form 3) Return all unused affected devices to the Recalling Fi
Quantity7 lots
DistributionU.S Nationwide Distribution.: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, VA, WA, and WI. O.U.S.: United Arab Emirates, France, Pakistan, Argentina, United Kingdom, Portugal, Austral
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K962495INDEFLATOR PLUS 30/PLUS 30 PRIORITY PACK9WITH THE.096 OR.115 ROTATING HEMOSTATIC VALVE)
Advanced Cardiovascular Systems, Inc., Sep 12, 1996