ClearanceBrief
Recall Z-0923-2022 ยท MAV Syringe, Balloon Inflation

Abbott Vascular

20/30 Priority Pack w/COPILOT - REF 1003327 WPL2122272-01 (2019-03-01)

DateMar 11, 2022
StatusOpen, Classified
ReasonDue to an increase in complaint trend for leaks and intermittent/loose connections.
Root causeComponent change control
ActionOn 03/11/2022, the firm sent an "URGENT FIELD SAFETY NOTICE/DEVICE RECALL" Letter via visit, email or mail, to customers informing them that specific lots of 20/30 INDEFLATOR, INDEFLATOR Plus 30 and associated Priority Packs my exhibit leaks and/or a loose connection at the rotating luer assembly or stopcock connection, which could lead to air ingress under vacuum. Customers are instructed to: 1)Immediately stop using the devices from the affected lots 2) Review their inventory, complete and return the provided Effectiveness Check Form 3) Return all unused affected devices to the Recalling Fi
Quantity131 lots
DistributionU.S Nationwide Distribution.: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, VA, WA, and WI. O.U.S.: United Arab Emirates, France, Pakistan, Argentina, United Kingdom, Portugal, Austral
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961471INDEFLATOR 20/30 INFLATION DEVICE AND 20/30 PRIORITY PACK (WITH THE .096 OR .115 ROTATING HEMOSTATIC VALVE)
Advanced Cardiovascular Systems, Inc., May 8, 1996