Recall Z-0926-2022 ยท FFL Dislodger, Stone, Basket, Ureteral, Metal

Philips North America, LLC

The IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for the following product model #s: 21075A, 21076A, 21078A, 989803162621, 989803162631, 989803162641.

DateFeb 7, 2022
StatusOpen, Classified
ReasonUpdate to instructions for use regarding the cleaning and disinfection process for the reusable probes.
Root causeDevice Design
ActionThe firm, Philips, issued "URGENT Medical Device Recall" Customer notification letters on 07 Feb 2022 via certified mail. Actions planned by the firm: An IFU addendum (part# 453564978601) was created that provides updated instructions for cleaning, disinfection, and sterilization. This addendum is to be downloaded www.philips.com/ifu and kept with the main IFU (part# 453564635891). Customer should take the following actions to prevent risk to patients or users: Prepare and soak in an enzymatic solution (detergent) to remove any debris prior to disinfection or sterilization. To reduce risk o
Quantity478,931 units
DistributionWorldwide distribution: US (nationwide) Including Puerto Rico and OUS (foreign) countries including: Afghanistan, Albania, Algeria, Angola, Argentina, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Da
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K962070YSI 400 SERIES ATUOCLAVABLE REUSABLE TEMPERATURE PROBE
Yellow Springs Instrument Co., Inc., Jul 19, 1996