Recall Z-0874-2022 ยท LLZ System, Image Processing, Radiological

Philips North America Llc

Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion (Model Numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228) running StentBoost Live (Model Number 459801781661) Software Version Number: R2.0

DateFeb 2, 2022
StatusOpen, Classified
ReasonDue to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
Root causeUnder Investigation by firm
ActionUrgent Medical Device Correction notification letters dated 2/2/22 were distributed to customers. 1. What is the problem and under what circumstances it can occur Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient. When the problem occurs, the area of the boosted images remains black, and the small X-ray viewer in the upper right corner will show a single static image of the previous run. The correct patien
Quantity338
DistributionGlobal Distribution. US Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141979ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES
Philips Medical Systems Netherland BV, Aug 19, 2014
K170144StentBoost Live
Philips Medical Stystems Nederland BV, Jun 7, 2017
K172822Azurion series R1.2
Philips Medical Systems Nederland B.V., Nov 22, 2017
K181830Azurion R2.0
Philips Medical Systems Nederland B.V., Aug 2, 2018