ClearanceBrief
Recall Z-0958-2022 ยท HWC Screw, Fixation, Bone

Acumed LLC

4.0mm x 10mm Cannulated Screw, Short Thread; Part number 3005-40010

DateMar 9, 2022
StatusCompleted
ReasonCannulated Screws 4.0mm x 10mm and 4.0mm x 12mm may not have sufficient axial pullout strength for the indicated use. Remote possibility of resulting in a mal or nonunion of the bone fragments, iatrogenic fracture, joint impingement or damage, or soft tissue damage.
Root causeDevice Design
ActionUrgent Medical Device recall notice dated 3/8/22 was send out to customers via mail. Special Instructions: IMMEDIATELY stop using and/or distributing the product and complete the following actions 1) Perform a physical count of your inventory. 2) Remove the affected product, such that they are out of service. 3) Record this data on the survey and Return the survey via email to RecallNotification@acumed.net. 4) To facilitate return of product. a) Independent Sales Consultants: contact Agent Inventory via email at agentinventory@acumed.net b) Domestic Direct Customers: contact Business Servic
Quantity537 units
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, Puerto Rico and the country of South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123890ACUMED CANNULATED SCREW SYSTEM
Acumed, LLC, Mar 18, 2013