Recall Z-0908-2022 ยท MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator

Respironics California, LLC

Philips Respironics V60 Plus Ventilator (All Models, including: 1138747, 1137276)

DateApr 18, 2022 (initiated Feb 28, 2022)
StatusOpen, Classified
ReasonVentilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
Root causeDevice Design
ActionOn March 14, 2022, Respironics California, LLC issued an Urgent Medical Device Correction notification to all affected consignees. On April 28, 2022 Respironics issued a revised customer communication titled Update: Urgent Medical Device Correction, dated April 20, 2022, which described mitigating actions. On June 30, 2022, Respironics issued an Updated Urgent Medical Device Correction Notification. This letter was to provide customers with an update on Philips Respironics actions to address an issue related to the internal source (35V Rail) that powers the V60/V60 Plus ventilators. In additi
Quantity12,145 units
DistributionWorldwide distribution US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of Afghanistan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102985V60 VENTILATOR WITH PPV AND AUTO-TRAK+ SOFTWARE OPTIONS
Respironics, Inc., Apr 5, 2011