Recall Z-0914-2022 ยท HWC Screw, Fixation, Bone

Nextremity Solutions

Nextra Hammertoe Correction System Driver SKU: NX-DR Orthopedic surgical instrument

DateFeb 10, 2022
StatusTerminated
ReasonProximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
Root causeNonconforming Material/Component
ActionNextremity Solutions issued Urgent Medical Device Recall letter on 2/10/22 via mail and email to Distributors and Risk Mangers. Nextremity Solutions products with the referenced part and lot number should not be used and should be quarantined and returned to Zimmer Biomet for return to Nextremity Solutions. If you have affected product at your facility, locate and quarantine affected product in your inventory. Zimmer Biomet Sales representative will remove the affected product from your facility. Complete Medical Device Recall Return Response and send to CorporateQuality.PostMarket@zim
Quantity239 units
DistributionDistribution US Nationwide, Japan, and the Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110445FLEXFUSION (TM) FIXATION IMPLANT
Nextremity Solutions, LLC, Mar 17, 2011