Recall Z-0931-2022 ยท NDW Assay, Porphyrin, Spectrophotometry, Lithium

Siemens Healthcare Diagnostics, Inc.

Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11097637

DateFeb 9, 2022
StatusOpen, Classified
ReasonSiemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.
Root causeDevice Design
ActionUrgent Medical Device Correction notification letters dated 2/9/22 were sent to customers. Actions to be Taken by the Customer Please review this letter with your Medical Director. Perform the instructions provided in the Additional Information section below. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical
Quantity2158
DistributionGlobal Distribution. US Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050374ADVIA IMS LITHIUM ASSAY
Bayer Healthcare, LLC, Jun 15, 2005