Recall Z-0928-2022 ยท CJY Azo Dye, Calcium

Siemens Healthcare Diagnostics, Inc.

Atellica CH Calcium_2 (CA_2), Siemens Material Number 11097644

DateFeb 9, 2022
StatusOpen, Classified
ReasonSiemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.
Root causeDevice Design
ActionUrgent Medical Device Correction notification letters dated 2/9/22 were sent to customers. Actions to be Taken by the Customer Please review this letter with your Medical Director. Perform the instructions provided in the Additional Information section below. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical
Quantity2,395
DistributionGlobal Distribution. US Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083386ADVIA CHEMISTRY CALCIUM_2 (CA_2) METHOD, MODELS P/M 02189699 (40 ML FILL), P/N 02189915 (70 ML FILL)
Siemens Healthcare Diagnostics, Apr 2, 2009