ClearanceBrief
Recall Z-0860-2022 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

1.5T and 3.0T SIGNA HDx, 1.5T and 3.0T SIGNA HDxt, and 1.5T SIGNA HDxt Mobile. Magnetic Resonance Imaging System

DateApr 4, 2022 (initiated Feb 18, 2022)
StatusOpen, Classified
ReasonThere is potential for the images to be flipped left to right.
Root causeDevice Design
ActionThis issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions a
Quantity653 units
DistributionWorldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, China, Denmark, Ecuador, Ethiopia, France, Germany, Greece, Hong Kong, India, Iran, Ireland, Italy, Japan, Korea, Kuwait, Lebanon, Mexico, Netherlands, New Zealand, Pakistan, Poland, Po
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052293GE 1.5T AND 3.0T SIGNA HDX MR SYSTEM
Ge Medical Systems, LLC, Sep 21, 2005
K1216761.5T SIGNA HDXT 3.0T SIGNA HDXT SIGNA VIBRANT
Ge Medical Systems, LLC, Sep 20, 2012