Recall Z-0857-2022 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

Signa 1.5T TwinSpeed Magnetic Resonance Imaging System

DateApr 4, 2022 (initiated Feb 18, 2022)
StatusOpen, Classified
ReasonThere is potential for the images to be flipped left to right.
Root causeDevice Design
ActionThis issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions a
Quantity10 units
DistributionWorldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, China, Denmark, Ecuador, Ethiopia, France, Germany, Greece, Hong Kong, India, Iran, Ireland, Italy, Japan, Korea, Kuwait, Lebanon, Mexico, Netherlands, New Zealand, Pakistan, Poland, Po
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0122001.5T INFINITY TWINSPEED MR SYSTEM