Recall Z-0837-2022 ยท LRS I.V. Start Kit

ROi CPS LLC

Regard Custom Procedure Kits containing BD Luer Lok Access Devices

DateFeb 24, 2022
StatusTerminated
ReasonRecalled devices were packaged in convenience kits.
Root causeMaterial/Component Contamination
ActionA customer letter dated February 24, 2022 was issued to the impacted consignee. The letter instructs the consignee to perform a microscopy to confirm an elevated RBC count in a urine sample as the impacted component may shed contaminant particles that are identified as RBCs when used in conjunction with the Sysmex UF-1000i or UX-2000 analyzers. A reply form should be completed and returned to ROi confirming receipt and understanding of the notification. End users should retain a copy of the BD notice and post near the analyzer as a visual reminder of the potential issue with the BD product.
Quantity1750 kits
DistributionUS Nationwide distribution in the state of Missouri.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991088VACUTAINER BRAND MULTIPLE SAMPLE LUER ADAPTER, BLOOD COLLECTION SETS AND SAFETY-LOK BLOOD COLLECTION, SAFETY BLOOD COLLE
Becton Dickinson Vacutainer Systems, May 28, 1999