ClearanceBrief
Recall Z-0871-2022 ยท CBK Ventilator, Continuous, Facility Use

Draeger Medical, Inc.

Babylog VN500 Ventilator with Software 2.51.01 and Lower- intended for the ventilation neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight Catalog Number: 8417400

DateFeb 14, 2022
StatusTerminated
ReasonCybersecurity Improvement Action-not equipped against potential cyber security threats
Root causeNonconforming Material/Component
ActionDraeger issued "Urgent Medical Device Correction Notices" on 2/14/22 to the Director, Clinical Engineering and Risk Manager via FedEx with tracking/confirmation of delivery. Letter states reason for recall, health risk and action to take: Continue to follow the recommendations in the instructions for use: o Limit or control physical access to the abovementioned ventilators. o Do not connect any non-approved devices to the USB, LAN and DVI interfaces. o Be attentive to notifications, alarms, and alerts Consider closing/covering all unused USB, LAN and DVI interfaces Should you decide to clo
Quantity13,687 (2,987 units US)
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093632INFINITY ACUTE CARE SYSTEM WORKSTATION NEONATAL CARE
Drager Medical AG & Co. Kgaa, Apr 5, 2010