ClearanceBrief
Recall Z-0917-2022 ยท DSA Cable, Transducer And Electrode, Patient, (Including Connector)

Volcano Corp

MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight Integrated (Part Number: 797403) IVUS systems

DateFeb 3, 2022
StatusOpen, Classified
ReasonMulti-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore, bottom protective beams will be replaced with full protective metal brackets.
Root causeDevice Design
ActionOn 02/03/22, correction notices were sent to customers who were asked to do the following: Notify all affected system users within the facility of this recall. The updated bracket improves protection on all sides of the MM-TSM touch screen that will prevent issues, cosmetic concerns, and maintain device integrity. If you are a distributor or have forwarded affected device(s) to another end user, it is imperative that all end-users with affected devices receive this Important Product Notice. Complete and return the customer response form. Customers with additional questions are encouraged to
Quantity1358
DistributionUS: AZ, IL, NV, MI, FL, NC, AK, PA, WI, MN, CA, AR, NJ, TX, OR, NM, SD, CO, KY, LA, DE, MA, MT, MO, VA, AL, NY, WV, SC, IN, GA, IA, OH, UT, NH, MS, TN, OK, ID, MD, HI, RI, WY, DC, WA, KS, NE, ND, CT, VT. OUS: SK, IT, DK, GB, DE, SE, IN, KW, QA, PL, NL, CL, JP, BE, AU, FR, PF, PT, NZ, CA, ES, PK, CY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.