Recall Z-0807-2022 ยท LIP Enzyme Linked Immunoabsorption Assay, Treponema Pallidum

Bio-Rad Laboratories, Inc.

BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL

DateFeb 8, 2022
StatusOpen, Classified
ReasonDue to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that consistently meet stability specifications
Root causeUnder Investigation by firm
ActionOn 02/08/2022, the firm sent out an "UGENT: Medical Device Correction" letter to customers via overnight FedEx informing them that the Recalling Firm continues to receive reports of false reactive RPR results despite labeling revisions and the previous Technical Bulleting dated December 2021 (LB003034revA). This is in response to ongoing concerns related to COVID-19 vaccine interference with RPR and the stability issue that led to the short dating of lot, the Recalling Firm is taking the additional field action to disable the RPR portion of the Syphilis Total and RPR panels until further noti
Quantity724 Calibrator sets (U.S.)
DistributionWorldwide distribution U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of France, Germany, Italy, Spain, Sa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170413BioPlex 2200 Syphilis Total & RPR, BioPlex 2200 Syphilis Total & RPR Calibrator Set, BioPlex 2200 Syphilis Total & RPR C
Bio-Rad Laboratories, May 11, 2017