Recall Z-0896-2022 ยท JKD Radioimmunoassay, C-Peptides Of Proinsulin

DRG International, Inc.

C-Peptide ELISA : IVD measurement of C-Peptide in serum, plasma (EDTA-, heparin- or citrate plasma) and urine. Catalog # EIA-1293

DateDec 21, 2021
StatusTerminated
ReasonODs of the standards too low(< 0.6 ODmax) may cause invalid runs, delayed patient results are possible
Root causeUnder Investigation by firm
ActionDRG issued letter via email on 12/28/21. Letter states reason for recall, health risk and Action to be taken by distributors and end-users: Please check your inventory for the respective lot, stop using these kits immediately and promptly quarantine these products. Only if you faced above mentioned issue of SO OD <1.2, please ask for replacement of the kits that were affected and provide your data. All unused kits of the affected lot will be replaced.Please reply immediately upon checking your stock/use to this notice to confirm. All replies should be received at once OR by January 7th, 2022.
Quantity3 Kits
DistributionNY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.