Recall Z-0885-2022 ยท OWB Interventional Fluoroscopic X-Ray System

Volcano Corporation

SyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.x

DateJan 19, 2022
StatusOpen, Classified
ReasonIf FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR Co-Registration results to be displayed, that may cause users to mistakenly use incorrect measurements, leading to inappropriate patient treatment. Image acquisition and processing system operators manual to be updated to include workflow alternatives.
Root causeDevice Design
ActionOn 01/19/22, correction notices were sent to customers who were informed of the following: If FFR measurement(s) are made prior to an iFR/FFR co-registration in the same procedural session. The user will be visually alerted with a warning message "Insufficient data, distal segment is not co-registered" on the display about the distal segment not being co-registered. The user may also notice the initial iFR/FFR co-registered distal value(s) will be higher than the correctly displayed distal iFR/FFR value(s). Recommended Course of Action: If prior to the iFR/FFR Pullback on the system, FFR mea
Quantity3,216
DistributionUS: WI, CA, CO, MO, ID, DC, LA, VA, KS, OK, GA, KY, IL, AZ, NY, IN, FL, TX, NC, OH, WA, OR, NM, MT, IA, MI, TN, MN, PA, AL, MA, WV, NV, NH, CT, NE, NJ, HI, AR, UT, MD, SC, MS, AK, DE OUS: JP, SA, GB, DE, PT, SG, PL, ZA, NL, MT, ES, IN, GR, EG, FI, HK, FR, NO, DK, BE, AT, IT, CO, KW, MX, PS, QA, CY,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132558SYNC-RX SYSTEM
Volcano Corporation, Dec 12, 2013
K151904SyncVision System
Volcano Corporation, Jul 31, 2015
K161756SyncVision System
Volcano Corporation, Sep 5, 2016
K172574SyncVision System
Volcano Corporation, Oct 5, 2017
K190626SyncVision System
Volcano Corporation, Apr 12, 2019