Recall Z-0836-2022 ยท OOI Real Time Nucleic Acid Amplification System

Meridian Bioscience Inc

Revogene, For In Vitro Diagnostic use in performing nucleic acid testing in clinical laboratories. Meridian branded Revogene: Catalog Number 610210 GenePOC branded Revogene: Catalog Number 130840

DateFeb 15, 2022
StatusOpen, Classified
ReasonThe photomultiplier tube, may have compromised functionality in some instruments. As a result, the user is likely to observe an increase in false-positive or indeterminate results.
Root causeComponent design/selection
ActionA Letter dated 02/15/2022 sent via FedEx notified users of the risk of false-positives and provided instructions on how to recognize a compromised photomultiplier tube via error codes and based on various combinations of positive and indeterminate results. If a positive result with the negative control is obtained or false positive suspected, users are to contact Technical Service and cease using the instrument. The company is developing software solution aimed at identifying photomultiplier tubes that are at risk for failure and is seeking a hardware fix to this issue.
Quantity452 units (398 US, 54 OUS)
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OR, PA, PR, RI, SD, TN, TX, UT, VA, VT, WI, WV and the countries of Australia, Chile, Italy, Switzerland.