Recall Z-0819-2022 · MEA Pump, Infusion, Pca

B Braun Medical Inc

PERFUSOR SPACE - US VERSION. Intended Use: syringe infusion pump that can be used to configure customized, tailor-made solutions Catalog Number: 8713030U

DateFeb 7, 2022
StatusTerminated
ReasonUnits not properly inspected prior to release to the market, may not function as indicated in the product labeling, resulting in syringe detection and infusion accuracy that can result in delays in therapy, overinfusion, or underinfusion.
Root causeUnder Investigation by firm
ActionBBraun issued "URGENT MEDICAL DEVICE CORRECTION" letter on 2/7/21 to End-Users. Letter states reason for recall, health risk and action to take: 1. Review the Device Correction Notification in its entirety and ensure that all users in your organization of theabove-mentioned product, and other concerned persons are informed about this voluntary correction. 2. Identify your current inventory of the affected serial (s) within your facility, cease use and quarantine product subject to correction. Do not destroy any affected product. 3. Utilizing the attached Product Correction Acknowledgement fo
Quantity6 units
DistributionUS Distribution to states of: MA, VT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092313PERFUSOR SPACE INFUSION SYRINGE PUMP SYSTEM
B.Braun Medical, Inc., Nov 4, 2009
K172831Perfusor® Space Syringe Infusion Pump System
B.Braun Medical, Inc., Jun 12, 2018
K191910SpaceStation MRI
B. Braun Melsugen AG, Mar 12, 2020