Recall Z-0855-2022 ยท OZP Automated Insulin Dosing Device System, Single Hormonal Control

Medtronic MiniMed

MiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892; MiniMed 620G (OUS Only): MMT-1750; MiniMed 640G (OUS Only): MMT-1711, MMT-1712, MMT-1751, MMT-1752; MiniMed 720G (OUS Only): MMT-1809, MMT-1810, MMT-1859, MMT-1860; MiniMed 740G

DateJan 27, 2022
StatusOpen, Classified
ReasonNew or replacement insulin pumps are not pre-programmed with basal rates or other verified settings (bolus wizard settings, sensor settings, etc.), which must be set up and saved on pumps prior to use. Patients who do not realize this will receive no basal insulin, which can lead to insulin under-delivery. Insulin under-delivery can cause severe hyperglycemia, which may lead to diabetic ketoacidosis. Current device labeling does not clearly identify that these insulin pumps are capable of functioning with the default basal setting of 0.0 units/hr for all 24 hours and that no error message will
Root causeLabeling design
ActionOn 01/27/22, correction notices were sent to patients, healthcare providers, and distributors who were asked to do the following: 1) Consult with your healthcare professional to determine the settings before using the pump. 2) Existing Users with replacement or upgrade devices should: - Verify current basal rate pump settings. See user guide Viewing your basal information section. Settings can be saved on CareLink or written down on a paper. - If the basal rate settings are not present on your pump, please take all the following actions: Locate pump settings and consult with your healthcare
Quantity175,801
DistributionUS: OH, UT, IL, NY, FL, LA, NE, NV, CA, VA, TX, DE, KS, MS, MI, AL, OK, AZ, SC, IN, AR, PA, WV, MA, RI, NJ, CO, TN, KY, GA, WI, MN, NM, MD, IA, NC, WY, MO, WA, OR, CT, ND, ME, VT, ID, SD, AK, MT, NH, DC, HI, PR, VI OUS: Australia, Japan, Korea, New Zealand, Taiwan, Viet Nam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.