Recall Z-0732-2022 ยท MUJ System, Planning, Radiation Therapy Treatment
RAYSEARCH LABORATORIES AB
RayStation/RayPlan- designed for treatment planning and analysis of radiation therapy. Software Version: 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 . Including Service Paks: Product name (build number) UDI-DI RayStation 6/RayPlan 2 (6.0.0.24) 07350002010013 RayStation 6/RayPlan
| Date | Dec 30, 2021 |
| Status | Open, Classified |
| Reason | If a new primary image set is selected while the cine loop is running, the primary image set will be displayed as both primary and secondary image set in all side-by-side views. This will also be true for any new patient or case opened while the cine loop is running |
| Root cause | Software design |
| Action | RaySearch issued Field Safety Notice, Medical Device Correction #90375 via E-mail, on 30 December 2021. Letter stated reason for recall, health risk and action to take: Do not change the primary image set while the cine loop is running. " If the primary image set has been changed while the cine loop is running, restart RayStation/RayPlan to restore correct behavior. " Educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the |
| Quantity | 75 Licenses |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K211867 | RayStation 11.0 RaySearch Laboratories AB (PUBL), Sep 8, 2021 |