ClearanceBrief
Recall Z-0869-2022 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Shimadzu Medical Systems

MODEL: X-RAY TV SYSSTEM SONIALVISION safire17

DateJan 10, 2022
StatusOpen, Classified
ReasonIt was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
Root causeSoftware design
ActionOn or about 1/7/2022, the firm sent an "Urgent: Voluntary Medical Device Recall Notice" to customers informing them that there is a potential issue with the adjustment procedure of the generator on the SonialVision (G$ and safire) Systems and FluoroSpeed X1 Systems. As a result of the Recalling Firm's investigation and testing it was proven that there is a problem with the adjustment procedure documented for the calibration of the system xray generator. This problem occurs during the calibration of the fluoroscopic mode which could cause x-ray radiation dose rate to exceed regulation and/or s
Quantity12 systems
DistributionU.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190373SonialVision G4
Shimadzu Corporation, Mar 15, 2019