Recall Z-0785-2022 ยท CBK Ventilator, Continuous, Facility Use

Philips Respironics, Inc.

Trilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.

DateJun 12, 2018
StatusOpen, Classified
ReasonIn 2018, Philips Respironics added foam replacement to the preventive maintenance schedule for these devices after discovering the polyester-based polyurethane (PE-PUR) foam may degrade into particles. This 2018 correction has been superseded by the device removal and rework described in Philips Respironics 2021 recall action for the same devices.
Root causeDevice Design
ActionIn 2018, Philips Respironics added foam replacement to the preventive maintenance schedule for these devices after discovering the polyester-based polyurethane (PE-PUR) foam may degrade into particles. This 2018 correction has been superseded by the device removal and rework described in Philips Respironics 2021 recall action for the same devices. A Field Communication for Trilogy Preventive Maintenance Update was distributed to customers 6/12/18. PRODUCT(S) AFFECTED All Trilogy models with Inlet Air Path Assembly PURPOSE / SCOPE This Field Communication announces: " The addition of the rep
Quantity229,353 devices
DistributionGlobal Distribution. US Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111610TRILOGY SERIES VENTILATOR WITH OXIMETRY
Respironics, Inc., Oct 14, 2011