Recall Z-0729-2022 ยท MAL Graft, Vascular, Synthetic/Biologic Composite

Maquet Cardiovascular, LLC

HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease. Product Code/REF Number: M002020952070

DateJan 14, 2022
StatusTerminated
ReasonOuter labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Straight Vascular Graft, D: 7 mm L: 60 cm). Mislabeled product could cause a disruption of the vascular anastomosis in the long term, creating a pseudo-aneurysm
Root causePackaging process control
ActionGetinge issued Urgent Medical Device Letter onJanuary 14, 2022, (via FedEx overnight) with adult signature required. Letter states reason for recall, health risk and action to take: Actions to be taken: " If you have already implanted the above-mentioned device, kindly notify us as soon as possible by contacting your Getinge representative or Customer Service (at 1-888-943-8872, options 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST. You will be contacted shortly by our Medical Affairs Officer in order to discuss the appropriate follow up. " Please examine your inv
Quantity1 unit
DistributionUS Nationwide distribution in the state of MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K954848HEMASHIELD MICROVEL VELOUR KNITTED VASCULAR GRAFT
Meadox Medicals, Div. Boston Scientific Corp., Feb 1, 1996