Recall Z-0736-2022 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Bolder Surgical, LLC

CoolSeal Generator, REF: CSL-200-50

DateJan 25, 2022
StatusTerminated
ReasonA potential manufacturing/assembly defect could result in the generator not meeting IEC 60601 standards for medical electrical equipment, and it can potentially, under specific conditions, lead to the delivery of unintended energy to the patient.
Root causeProcess control
ActionOn 01/25/22, recall notices were hand delivered to U.S. customers and emailed to O.U.S. customers. Customers were asked to do the following: Discontinue use and return affected devices to the recalling firm, and complete and return the acknowledgment form. Customers with additional questions were encouraged to contact the recalling firm at 1-866-683-1743.
Quantity10
DistributionUS: NY, MO, DE, PA, FL OUS: Italy, Chile, UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K202114CoolSeal Generator/CSL-200-50, CoolSeal Trinity (30 cm shaft)/CSL-TR105-30, CoolSeal Trinity (37 cm shaft)/CSL-TR105-37,
Bolder Surgical, LLC, Oct 6, 2020