Recall Z-0742-2022 ยท LKQ Antibody Igm,If, Cytomegalovirus Virus

bioMerieux, Inc.

VIDAS CMV IgM, 30 tests, REF 30205-01, IVD, Rx.

DateMar 14, 2022 (initiated Feb 9, 2022)
StatusTerminated
ReasonInvalid calibrations (out of range high) while using the product.
Root causeNonconforming Material/Component
ActionThe recalling firm issued letters dated 2/9/2022 via Fed Ex Express Priority Overnight delivery service to their U.S. consignees. The letter described the issue, the impact to the user, and the required actions. Those actions included instructing the consignee to stop using and destroy any stock of the recalled lot and to complete the Acknowledgement Form and return it via email or FAX. The quantity discarded was to be completed on the Acknowledgement Form.
Quantity480 kits
DistributionDistribution was made to CA, FL, GA, IL, IN, LA, MA, MD, ME, MO, NJ, NY, OH, PA, TX, VA, WI, and WV. There was government distribution but no military or foreign distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K933549VIDAS CMV IGM ASSAY
Biomerieux Vitek, Inc., Aug 2, 1994